Influence factors of Czerny-Turner edge band astigmatism correction
(2.Opto-electronics Engineer Institute,Changchun University of Science and Technology, Changchun ,China 130022)
【Abstract】To improve the weak correction capacity of a wide band Czerny-Turner structure, the influence factors on the astigmatism correction of an optical system were analyzed. The Matlab software was used to simulate and analyze the reasons that the astigmatism was produced and the traditional method could not suppress the astigmatism. Then, the dependence of the angle difference α between the off-axis angle of collimating mirror and the off-axis angle of focusing mirror on the astigmatism S of the optical system was discussed. The astigmatism correction of wide band Czerny-Turner structure was simulated at different α values. To verify the validity of the theoretical analysis, an astigmatic Czerny-Turner system ranging from 900 to 1 700 nm was designed, the Zemax was used to perform the ray tracing and optimal design and the design results were processed and analyzed. The simulation results demonstrate that the short-wave band astigmatism correction ability is getting stronger with increasing the angle difference, which may achieve about 1. 6 times on increase. However, long-wave band astigmatism correction ability is weaker and weaker, which may reduce about 1.27 times on average. It suggests that the reasonable selection of the angle difference can provide the theoretical guidance for correcting the astigmatism of wide band Czerny-Turner structures.
【Keywords】 optical design; Czerny-Turner structure; astigmatism correction; edge band;
【DOI】
【Funds】
(Translated by caizhijian)
-
References
[1]CHEN W, ZHENG Y Q, XUE Q SH. Airborne imaging spectrometer with wide field of view and large relative-aperture[J]. Opt. Precision Eng., 2015, 23(1): 15-21. (in Chinese)
[2]ZHANG J, WANG SH R, HUANG Y, et al. Status and development of limb imaging spectrometer[J]. Chinese Optics, 2013, 6(5): 692-700. (in Chinese)
[3]CUI J CH, TANG Y G, HAN P P, et al. Development of diagnostic imaging spectrometer for tumor on-line operation[J]. Opt. Precision Eng., 2013, 21(12): 3043-3049. (in Chinese)
[4]CAO H X, WU N, FENG SH L, et al. Crossspectral calibration for monochromator and imaging spectrometer[J]. Opt. Precision Eng., 2014, 22(10): 2586-2591. (in Chinese)
[5]MASAYUKI F, TAKEHIKO TAKENOUCHI, KEI-ICHI K. Highly efficient and aberration-corrected spectrometer for advanced Raman spectroscope[J]. Applied Optics, 2002, 41(22): 4655-4665.
[6]XU L, CHEN K X, HE Q SH, et al. Design of freeform mirrors in Czerny-Turner spectrometers to suppress astigmatism[J]. Applied Optics, 2009, 48(15): 2871-2879.
[7]SUNG K, THOMPSON K P, ROLLAND J. Broadband antigmatism-corrected Czerny-Turner spectrometer[J]. Optics Experess, 2010, 18(22): 23378-23384.
[8]QU B X, GUO X, YU F H. Optimization of free-form cylindrical lens in Czerny-Turner spectrometer system to correct the astigmastism by using user defined error function[J]. SPIE, 2012, 8416: 84161I.
[9]XUE Q SH. Astigmastism-corrected Czerny-Turner imaging spectrometer for broadband spectral simultaneity[J]. Applied Optics, 2011, 50(10): 1338-1344.
[10]TANG Y, ZHANG ZH G, CHEN Y A, et al. Designed of visible-ultraviolet cylinder mirror imaging spectrometer[J]. Acta Optica Sinica, 2013, 33 (3): 0330004. (in Chinese)
[11]AUSTIN D R, WITTING T, WALMSLEY I A. Broadband astigmatism-free Czerny-Turner imaging spectrometer using spherical mirrors[J]. Applied Optics, 2009, 48(19): 3846-3853.
[12]BATE B, DOWELL M, NEWTON A C. Correction of astigmatism in a Czerny-Turner spectrograph using aplane grating in divergent illumination[J]. J. Phys., 1970, E3(3): 206-210.
ISSN:1004-924X
CN: 22-1198/TH
Vol 24, No. 10, Pages 2384-2391
October 2016
Downloads:0